Description
General
| Weight | 10.6 lbs. without paper and battery 11.7 lbs. with paper and battery |
| Dimensions | 8.9” x 10.4” x 7.9” With soft handle: 8.9” x 8.75” x 7.9” |
Display
| Display Type | Bright, Color, LCD TFT |
| Brightness | User adjustable |
| Resolution | 640 x 480 pixels |
| Waveform Traces | Up to 4 waveforms |
| Display Modes | Standard, High Contrast, NVG (night vision goggle) Friendly |
| Waveform Display Width | 11.5 cm |
Power
| Type | Rechargeable, user replaceable, lithium-ion battery |
| Operating Time | 6 hours (display brightness at 30%, ECG, SpO2, EtCO2, IPx3, Temp x2, and NIBP every 15 min, and 10 × 200-joule shocks) |
| Battery Size | 182 mm x 95 mm x 31 mm |
| Battery Weight | 1.36 lbs |
| Battery Indicators | 5 battery capacity LED indicators, fault indicator, recalibration indicator |
| Recharge Rate | 4 hours to 100% |
| AC Power Adapter | 100–240 VAC 50, 60, 400 Hz (@110V) |
Environmental
| Operating Temperature | 0° – 50° C |
| Operating Relative Humidity | 15% – 95%, non-condensing |
| Operating Altitude | Up to 15,000 ft |
| Transport/Storage Temperature | -30° – 70° C |
| Shock | IEC 60068-2-27, 100g, 6ms half sine |
Temperature
| Channels | Two YSI 400/700 compatible channels |
| Range | 0° – 50° C |
| Units | °C or °F |
| Display | T1, T2, and Delta temp |
| Accuracy | ± 0.1° C |
CPR Feedback Technology
| See-Thru CPR® | Artifact filtering |
| Real CPR Help® | Real-time depth and rate feedback |
| CPR Dashboard™ | Numeric readout of depth and rate, release indicator, perfusion performance indicator (PPI) |
Defibrillator
| Waveform | Rectilinear Biphasic™ |
| Energy Selections | 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 70, 85, 100, 120, 150, 200 joules |
| Energy Display | Monitor display indicates both selected and delivered energy |
| Charge Time | < 7 seconds to 200 J, typical |
| Charge Controls | Control on apex paddle and on device front panel |
| Paddles | External anterior/anterior adult and pediatric. Adult paddles slide off to expose pediatric paddles. Compatible with ZOLL external paddles (8000-1010-01), ZOLL internal paddles, and the Propaq MD multi-function therapy cable. |
| Synchronized Mode | Synchronizes defibrillator pulse to patient’s R-wave. “SYNC” message displayed on monitor. Marker on display and recorder paper identifies R-wave discharge point. |
| Built-In Defibrillator Tester | Tests defibrillator energy output and continuity of universal cable and paddles |
| Modes | AED function with manual override, manual defibrillation |
Invasive Pressure
| Channels | Three channels |
| Pressure Range | -30 – 300 mmHg |
| Pulse Rate Measurement Range | 25 – 250 BPM |
| Formats | S/D, S/D (M), (M) user selectable |
| User-Selectable Labels | P1, P2, P3, ABP, AO, ART, CVP, BAP, FAP, LAP, PAP, RAP, UAP, UVP, ICP |
| Transducer Requirements | 5 μV/V/mmHg |
| Zero Adjustment | ± 200 mmHg |
| Numeric Pressure Accuracy | ± 2 mmHg or 2% of reading, plus transducer error |
| Transducer Connector | Standard 6-pin AAMI |













Reviews
There are no reviews yet.